10–20 system (EEG)

Method of scalp electrodes placement

  • 6 min read
  • 3 chapters

Related on Lably

Through Electrode and Scientific method

Across all phases of consciousness, brains produce different, objectively recognizable and distinguishable electrical patterns, which can be detected by electrodes on the skin. These patterns vary, and are affected by multiple extrinsic factors, including age, prescription drugs, somatic diagnoses, history of neurologic insults/injury/trauma, and substance abuse.

The "10" and "20" refer to the fact that the actual distances between adjacent electrodes are either 10% or 20% of the total front–back or right–left distance of the skull. For example, a measurement is taken across the top of the head, from the nasion to inion. Most other common measurements ('landmarking methods') start at one ear and end at the other, normally over the top of the head. Specific anatomical locations of the ear used include the tragus, the auricle and the mastoid.

Electrode labeling

Each electrode placement site has a letter to identify the lobe, or area of the brain it is reading from: pre-frontal (Fp), frontal (F), temporal (T), parietal (P), occipital (O), and central (C). Note that there is no "central lobe"; due to their placement, and depending on the individual, the "C" electrodes can exhibit/represent EEG activity more typical of frontal, temporal, and some parietal-occipital activity, and are always utilized in polysomnography sleep studies for the purpose of determining stages of sleep.

There are also (Z) sites: A "Z" (zero) refers to an electrode placed on the midline sagittal plane of the skull, (FpZ, Fz, Cz, Oz) and is present mostly for reference/measurement points. These electrodes will not necessarily reflect or amplify lateral hemispheric cortical activity as they are placed over the corpus callosum, and do not represent either hemisphere adequately. "Z" electrodes are often utilized as 'grounds' or 'references,' especially in polysomnography sleep studies, and diagnostic/clinical EEG montages meant to represent/diagnose epileptiform seizure activity, or possible clinical brain death. The required number of EEG electrodes, and their careful, measured placement, increases with each clinical requirement and modality.

Even-numbered electrodes (2,4,6,8) refer to electrode placement on the right side of the head, whereas odd numbers (1,3,5,7) refer to those on the left; this applies to both EEG and EOG (electrooculogram measurements of eyes) electrodes, as well as ECG (electrocardiography measurements of the heart) electrode placement. Chin, or EMG (electromyogram) electrodes are more commonly just referred to with "right," "left," and "reference," or "common," as there are usually only three placed, and they can be differentially referenced from the EEG and EOG reference sites.

The "A" (sometimes referred to as "M" for mastoid process) refers to the prominent bone process usually found just behind the outer ear (less prominent in children and some adults). In basic polysomnography, F3, F4, Fz, Cz, C3, C4, O1, O2, A1, A2 (M1, M2), are used. Cz and Fz are 'ground' or 'common' reference points for all EEG and EOG electrodes, and A1-A2 are used for contralateral referencing of all EEG electrodes. This EEG montage may be extended to utilize T3-T4, P3-P4, as well as others, if an extended or "seizure montage" is called for.