Collimated transmission theory

  • 3 min read
  • 3 chapters

Related on Lably

Through Hole, Photon and Refractive index

In the collimated transmission method, a laser beam is directed perpendicularly to the material and the detection of reemitted light gives information about the total interactive effect of the optical properties of the material. The use of multiple wavelengths can produce a spectra with more detailed information about the composition of the tissue or material (spectroscopy). While this method is simple and requires only minimal instrumentation, it produces error related to multiple scattering events and specular reflection. Some useful equations governing the properties include:

μa = absorption coefficient = Na𝜎a; μs = scattering coefficient = Ns𝜎s; g = scattering anisotropy = <cos(𝛳)>; μt = extinction coefficient = μa + μs

Where Na is the number of absorbers in a medium, 𝜎a is the absorption cross section, Ns is the number of scatterers in a medium, 𝜎s is the scattering cross section, and 𝛳 is the scattering angle.

History

The collimated transmission method has been used to measure the optical properties of biological tissues since the early 1980s. A collimated light source was generated by a laser or with a diffuse source and a collimator. Unscattered light transmission through the tissue was detected and Beer's law was used to estimate the extinction coefficient μt. This was done for human, porcine, rodent, bovine, and chicken tissues and compared with theoretical models to predict reliable optical properties. Knowledge of these properties were crucial for dosimetry studies.

Additionally, this method has been applied to measuring optical properties of turbid media for improving the simulation of tissue phantoms.

By the late 1980s, several studies also tested optical properties of biological tissues at different wavelengths to produce spectra.