Electron capture ionization

  • 4 min read
  • 5 chapters

Related on Lably

Through Atom, Molecule and Electron

A+e−→MA−

where the M over the arrow denotes that to conserve energy and momentum a third body is required (the molecularity of the reaction is three).

Electron capture can be used in conjunction with chemical ionization.

Electron-capture mass spectrometry

Electron-capture mass spectrometry (EC-MS) is a type of mass spectrometry that uses electron capture ionization to form negative ions from chemical compounds with positive electron affinities. The approach is particularly effective for electrophiles. In contrast to electron ionization, EC-MS uses low energy electrons in a gas discharge. EC-MS will cause less fragmentation of molecules compared to electron ionization.

Negative ion formation

Resonance electron capture

Resonance electron capture is also known as nondissociative EC. The compound captures an electron to form a radical anion. The energy of the electrons are about 0 eV. The electrons can be created in the Electron Ionization source with moderating gas such as H2, CH4, i-C4H10, NH3, N2, and Ar. After the ion captures the electron, the complex formed can stabilize during collisions and produce a stable anion that can be detected in a mass spectrometer.

AB + e− → AB−•

Dissociative resonance capture

In dissociative resonance capture, the compound fragments resulting in electron capture dissociation (ECD). ECD forms an anion fragment and a radical fragment. The energy of the electrons are from 0-15 eV, but the optimum energy can vary depending on the compound.

AB+e−⟶A−+B∙

Ion-pair formation

With electrons of energy greater than 10 ev, negative ions can also be formed through ion-pair formation.

AB + e− → A− + B+ + e−

Calibration of the mass spectrometer is important in electron capture ionization mode. A calibration compound is needed to ensure reproducibility in EC-MS. It is used to ensure that the mass scale used is correct and that the groups of ions are constant on a regular basis.

Fragmentation in ECI has been studied by tandem mass spectrometry.

The technique can be used with gas chromatography-mass spectrometry.