Fluid-attenuated inversion recovery
Medical imaging technique
- 1 min read
- 2 chapters
Explore on Lably
ToursExplore Lably’s 3D toursTechnique
By carefully choosing the inversion time (TI), the signal from any particular tissue can be nulled. The appropriate TI depends on the tissue via the formula:
In other words, one should typically use a TI of around 70% of the T1 value. In the case of CSF suppression, the long TI (inversion time) is adjusted to a zero crossing point for water (none of its signal is visible), so the signal of the CSF is theoretically "erased" from the image.
The standard T2 variant of FLAIR uses a long TE to develop T2-weighting and a long repetition time (TR) to increase signal to noise ratio (SNR), while the T1 variant of FLAIR uses a short echo time (TE) and short TR to develop T1-weighting. The TR also controls the SNR efficiency of the scan, which is optimal only for a particular T1.
Sign in to keep reading
One more chapter is free with a Lably account.
Clinical applications
The FLAIR sequence analysis has been especially useful in the evaluation and study of CNS disorders, involving:
- Lacunar infarction
- Multiple sclerosis (MS) plaques
- Subarachnoid haemorrhage
- Head trauma
- Meningitis and other leptomeningeal diseases