Impedance microbiology

Measuring the microbial density in a sample via its electrical parameters

  • 5 min read
  • 3 chapters

Related on Lably

Through Microbiology, Bacteria and Microbial metabolism

Principle of operation

When a pair of electrodes are immersed in the growth medium, the system composed of electrodes and electrolyte can be modeled with the electrical circuit of Fig. 1, where Rm and Cm are the resistance and capacitance of the bulk medium, while Ri and Ci are the resistance and capacitance of the electrode-electrolyte interface. However, when frequency of the sinusoidal test signal applied to the electrodes is relatively low (lower than 1 MHz), the bulk capacitance Cm can be neglected, and the system can be modeled with a simpler circuit consisting only of a resistance Rs and a capacitance Cs in series. The resistance Rs accounts for the electrical conductivity of the bulk medium, while the capacitance Cs is due to the capacitive double-layer at the electrode-electrolyte interface. During the growth phase, bacterial metabolism transforms uncharged or weakly charged compounds of the bulk medium into highly charged compounds that change the electrical properties of the medium. This results in a decrease of resistance Rs and an increase of capacitance Cs.

In impedance microbiology, the technique works this way. The sample with the initial unknown bacterial concentration (C0) is placed at a temperature favoring bacterial growth (in the range of 37 to 42 °C if a mesophilic microbial population is the target), and the electrical parameters Rs and Cs are measured at regular time intervals of a few minutes by means of a couple of electrodes in direct contact with the sample.

Until the bacterial concentration is lower than a critical threshold, CTH the electrical parameters Rs and Cs remain essentially constant (at their baseline values). CTH depends on various parameters such as electrode geometry, bacterial strain, chemical composition of the growth medium, etc., but it is always in the range of 106 to 107 cfu/ml.

When the bacterial concentration increases over CTH, the electrical parameters deviate from their baseline values (generally in the case of bacteria there is a decrease of Rs and an increase of Cs; the opposite happens in the case of yeasts).

The time needed for the electrical parameters Rs and Cs to deviate from their baseline value is referred to as Detect Time (DT) and is the parameter used to estimate the initial unknown bacterial concentration C0.

In Fig. 2 a typical curve for Rs as well as the corresponding bacterial concentrations are plotted vs. time. Fig. 3 shows typical Rs curves vs time for samples characterized by different bacterial concentration. Since DT is the time needed for the bacterial concentration to grow from the initial value C0 to CTH, highly contaminated samples are characterized by lower values of DT than samples with low bacterial concentration. Given C1, C2 and C3 the bacterial concentrations of three samples with C1 > C2 > C3, it is DT1 < DT2 < DT3. Data from literature show how DT is a linear function of the logarithm of C0:

DT=A⋅log10⁡(C0)+B

where the parameters A and B are dependent on the particular type of samples under test, the bacterial strains, the type of enriching medium used, and so on. These parameters can be calculated by calibrating the system using a set of samples whose bacterial concentration is known and calculating the linear regression line that will be used to estimate the bacterial concentration from the measured DT.

Impedance microbiology has different advantages over the standard plate count technique to measure bacterial concentration. It is characterized by faster response time. In the case of mesophilic bacteria, the response times range from 2 – 3 hours for highly contaminated samples (105 - 106 cfu/ml) to over 10 hours for samples with very low bacterial concentration (less than 10 cfu/ml). As a comparison, for the same bacterial strains, the Plate Count technique is characterized by response times from 48 to 72 hours.

Impedance microbiology is a method that can be easily automated and implemented as part of an industrial machine or realized as an embedded portable sensor, while plate count is a manual method that needs to be carried out in a laboratory by trained personnel.