Mycobacteria growth indicator tube

Bacteriology medium

  • 2 min read
  • 2 chapters

Related on Lably

Through Antibiotic, Blood and Urine

All types of clinical specimens, pulmonary as well as extra-pulmonary (except blood and urine), can be processed for primary isolation in the MGIT tube using conventional methods. After processed specimen is inoculated, MGIT tube must be continuously monitored either manually or by automated instruments until positive or the end of the testing protocol.

Principles of the procedure

A fluorescent compound is embedded in silicone on the bottom of 16 × 100 mm round bottom tubes. The fluorescent compound is sensitive to the presence of oxygen dissolved in the broth. Initially, the large amount of dissolved oxygen quenches emissions from the compound and little fluorescence can be detected. Later, actively respiring microorganisms consume the oxygen and allow the fluorescence to be detected.

Tubes are filled with samples in the broth and continuously incubated at 37 °C. The tubes are monitored for increasing fluorescence to determine if the tube is instrument positive; i.e., the test sample contains viable organisms. Fluorescence can be recorded by automated instruments such as Becton Dickinson's BACTEC MGIT 960 System, or manually using the BD BACTEC microMGIT fluorescence reader or a Wood's lamp or other long-wave UV light source.