Quadrupole mass analyzer

  • 3 min read
  • 3 chapters

Related on Lably

Through Mass, Electric field and Hyperbola

Principle of operation

The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically, and a radio frequency (RF) voltage with a DC offset voltage is applied between one pair of rods and the other. Ions travel down the quadrupole between the rods. Only ions of a certain mass-to-charge ratio will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This permits selection of an ion with a particular m/z or allows the operator to scan for a range of m/z-values by continuously varying the applied voltage. Mathematically this can be modeled with the help of the Mathieu differential equation.

Ideally, the rods are hyperbolic, however cylindrical rods with a specific ratio of rod diameter-to-spacing provide an easier-to-manufacture adequate approximation to hyperbolas. Small variations in the ratio have large effects on resolution and peak shape. Different manufacturers choose slightly different ratios to fine-tune operating characteristics in context of anticipated application requirements. Since the 1980s, the MAT company and subsequently Finnigan Instrument Corporation used hyperbolic rods produced with a mechanical tolerance of 0.001 mm, whose exact production process was a well-kept secret within the company.