Conductivity (electrolytic)

Measure of the ability of a solution containing electrolytes to conduct electricity

  • 11 min read
  • 6 chapters

Related on Lably

Through Electrolyte, Electrical resistivity and conductivity and International System of Units

Conductivity measurements are used routinely in many industrial and environmental applications as a fast, inexpensive and reliable way of measuring the ionic content in a solution. For example, the measurement of product conductivity is a typical way to monitor and continuously trend the performance of water purification systems.

In many cases, conductivity is linked directly to the total dissolved solids (TDS).

High-quality deionized water has a conductivity of

κ=0.05501±0.0001 μS/cm at 25 °C.

This corresponds to a specific resistivity of

ρ=18.18±0.03 MΩ⋅cm.

The preparation of salt solutions often takes place in unsealed beakers. In this case the conductivity of purified water often is 10 to 20 times higher. A discussion can be found below.

Typical drinking water is in the range of 200–800 μS/cm, while sea water is about 50 mS/cm (or 0.05 S/cm).

Electrolytic conductivity varies from about 10E-10 S/m for purified toluene up to about 10 S/m for recently discovered highly concentrated “water-in-salt” solutions.

Conductivity of aqueous and other polar solutions is traditionally determined by connecting the electrolyte in a Wheatstone bridge. Dilute solutions follow Kohlrausch's law of concentration dependence and additivity of ionic contributions. Lars Onsager gave a theoretical explanation of Kohlrausch's law by extending Debye–Hückel theory.

Conductivity of low- and non-polar solutions is very low. It is also associated with motion of ions in electric field, but the nature and solvation of ions are different, as described on the page conductivity (non-aqueous). It is usually measured with probes having low cell constant and applying low frequency electric field.

Notation is treacherous in this domain. Rho (ρ) reliably denotes specific resistivity, usually quantified in units of ohm-meter. Kappa and sigma (κ and σ) are inversely proportional to sigma, with kappa preferred in classical electrochemistry and sigma preferred in Maxwellian electrodynamics. Unfortunately, both kappa and sigma have prominent confounds that can trip up the unwary. In electrochemistry, kappa is also used to denote inverse Debye screening length. In electrodynamics, sigma is also used to denote double-layer charge density.

Units

The SI unit of conductivity is S/m, and unless otherwise qualified, it refers to 25 °C. More generally encountered is the traditional unit of μS/cm.

The commonly used standard cell has a width of 1 cm, and thus for very pure water in equilibrium with air would have a resistance of about 106 ohms, known as a megohm. Ultra-pure water could achieve 18 megohms or more. Thus in the past, megohm-cm was used, sometimes abbreviated to "megohm". Sometimes, conductivity is given in "microsiemens" (omitting the distance term in the unit). While this is an error, it can often be assumed to be equal to the traditional μS/cm. Often, by typographic limitations μS/cm is expressed as uS/cm.

The conversion of conductivity (in μS/cm) to the total dissolved solids (in mg/kg) depends on the chemical composition of the sample and can vary between 0.54 and 0.96. Typically, the conversion is done assuming that the solid is sodium chloride; 1 μS/cm is then equivalent to about 0.64 mg of NaCl per kg of water.

Molar conductivity has the SI unit S⋅m2⋅mol−1. Older publications use the unit Ω−1⋅cm2⋅mol−1.