The fluorescence lifetime (FLT) of the fluorophore, rather than its intensity, is used to create the image in FLIM. Fluorescence lifetime depends on the local micro-environment of the fluorophore, thus precluding any erroneous measurements in fluorescence intensity due to change in brightness of the light source, background light intensity or limited photo-bleaching. This technique also has the advantage of minimizing the effect of photon scattering in thick layers of sample. Being dependent on the micro-environment, lifetime measurements have been used as an indicator for pH, viscosity and chemical species concentration.

Fluorescence lifetimes

A fluorophore which is excited by a photon will drop to the ground state with a certain probability based on the decay rates through a number of different (radiative and/or nonradiative) decay pathways. To observe fluorescence, one of these pathways must be by spontaneous emission of a photon. In the ensemble description, the fluorescence emitted will decay with time according to

I(t)=I0e−t/τ

where

1τ=∑ki.

In the above, t is time, τ is the fluorescence lifetime, I0 is the initial fluorescence at t=0, and ki are the rates for each decay pathway, at least one of which must be the fluorescence decay rate kf. More importantly, the lifetime, τ is independent of the initial intensity and of the emitted light. This can be utilized for making non-intensity based measurements in chemical sensing.

Sign in to keep reading

3 more chapters are free with a Lably account.

More in Human body