Lably
Skip to content

Light sheet fluorescence microscopy

Fluorescence microscopy technique

  • 11 min read
  • 4 chapters

Related on Lably

Through Cell biology and Refractive index

This method is used in cell biology and for microscopy of intact, often chemically cleared, organs, embryos, and organisms.

Starting in 1994, light-sheet fluorescence microscopy was developed as orthogonal plane fluorescence optical sectioning microscopy or tomography (OPFOS) mainly for large samples and later as the selective/single plane illumination microscopy (SPIM) also with sub-cellular resolution. This introduced an illumination scheme into fluorescence microscopy, which has already been used successfully for dark-field microscopy under the name ultramicroscopy.

Setup

Basic setup

In this type of microscopy, the illumination is done perpendicularly to the direction of observation (see schematic image at the top of the article). The expanded beam of a laser is focused in only one direction by a cylindrical lens, or by a combination of a cylindrical lens and a microscope objective as the latter is available in better optical quality and with higher numerical aperture than the first. This way a thin sheet of light or light-sheet is created in the focal region that can be used to excite fluorescence only in a thin slice (usually a few micrometers thin) of the sample.

The fluorescence light emitted from the light-sheet is then collected perpendicularly with a standard microscope objective and projected onto an imaging sensor (usually a CCD, electron-multiplying CCD or CMOS camera). In order to let enough space for the excitation optics/light-sheet an observation objective with high working distance is used. In most light-sheet fluorescence microscopes the detection objective and sometimes also the excitation objective are fully immersed in the sample buffer, so usually the sample and excitation/detection optics are embedded into a buffer-filled sample chamber, which can also be used to control the environmental conditions (temperature, carbon dioxide level ...) during the measurement. The sample mounting in light-sheet fluorescence microscopy is described below in more detail.

As both the excitation light-sheet and the focal plane of the detection optics have to coincide to form an image, focusing different parts of the sample can not be done by translating the detection objective, but usually the whole sample is translated and rotated instead.

Extensions of the basic idea

In recent years, several extensions to this scheme have been developed:

  • The use of two counter-propagating light-sheets helps to reduce typical selective plane illumination microscopy artifacts, like shadowing (see first z-stack above)
  • In addition to counter-propagating light-sheets a setup with detection from two opposing sides has been proposed in 2012. This allows measurement of z- and rotation-stacks for a full 3D reconstruction of the sample more rapidly.
  • The light-sheet can also be created by scanning a normal laser focus up and down. This also allows use of self-reconstructing beams (such as bessel beams or Airy beams) for the illumination which improve the penetration of the light-sheet into thick samples, as the negative effect of scattering on the light-sheet is reduced. These self-reconstructing beams can be modified to counteract intensity losses using attenuation-compensation techniques, further increasing the signal collected from within thick samples.
  • In oblique plane microscopy (OPM) the detection objective is used to also create the light-sheet: The light-sheet is now emitted from this objective under an angle of about 60°. Additional optics is used to also tilt the focal plane used for detection by the same angle.
  • Light-sheet fluorescence microscopy has also been combined with two-photon (2P) excitation, which improves the penetration into thick and scattering samples. Use of 2P excitation in near-infrared wavelengths has been used to replace 1P excitation in blue-visible wavelengths in brain imaging experiments involving response to visual stimuli.
  • Selective plane illumination microscopy can also be combined with techniques such as fluorescence correlation spectroscopy, to allow spatially resolved mobility measurements of fluorescing particles (e.g. fluorescent beads, quantum dots or fluorescently labeled proteins) inside living biological samples.
  • Also a combination of a selective plane illumination microscope with a gated image intensifier camera has been reported that allowed measuring a map of fluorescence lifetimes (fluorescence lifetime imaging, FLIM).
  • Light-sheet fluorescence microscopy was combined with super-resolution microscopy techniques to improve its resolution beyond the Abbe limit. Also a combination of stimulated emission depletion microscopy (STED) and selective plane illumination microscopy has been published, that leads to a reduced light-sheet thickness due to the stimulated emission depletion microscopy effect. See also the section on the power of resolution of light-sheet fluorescence microscopy below.
  • Light-sheet fluorescence microscopy was modified to be compatible with all objectives, even coverslip-based, oil-immersion objectives with high numerical aperture to increase native spatial resolution and fluorescence detection efficiency. This technique involves tilting the light-sheet relative to the detection objective at a precise angle to allow the light-sheet to form on the surface of glass coverslips.
  • Light-sheet fluorescence microscopy was combined with Adaptive Optics techniques in 2012 to improve the depth of imaging in thick and inhomogenous samples at a depth of 350 um. A Shack Hartmann wavefront sensor was positioned in the detection path and guide stars are used in a close feedback loop. In his thesis, the author discuss the advantage of having Adaptive Optics both in the illumination and detection path of the light-sheet fluorescence microscope to correct aberrations induced by the sample.

Sample mounting

The separation of the illumination and detection beampaths in light-sheet fluorescence microscopy (except in oblique plane microscopy) creates a need for specialized sample mounting methods. To date most light-sheet fluorescence microscopes are built in such a way that the illumination and detection beampath lie in a horizontal plane (see illustrations above), thus the sample is usually hanging from the top into the sample chamber or is resting on a vertical support inside the sample chamber. Several methods have been developed to mount all sorts of samples:

  • Fixed (and potentially also cleared) samples can be glued to a simple support or holder and can stay in their fixing solution during imaging.
  • Larger living organisms are usually sedated and mounted in a soft gel cylinder that is extruded from a (glass or plastic) capillary hanging from above into the sample chamber.
  • Adherent cells can be grown on small glass plates that are hanging in the sample chamber.
  • Plants can be grown in clear gels containing a growth medium. The gels are cut away at the position of imaging, so they do not reduce the light-sheet and image quality by scattering and absorption.
  • Liquid samples (e.g. for fluorescence correlation spectroscopy) can be mounted in small bags made of thin plastic foil matching the refractive index of the surrounding immersion medium in the sample chamber.

Some light-sheet fluorescence microscopes have been developed where the sample is mounted as in standard microscopy (e.g. cells grow horizontally on the bottom of a petri dish) and the excitation and detection optics are constructed in an upright plane from above. This also allows combining a light-sheet fluorescence microscope with a standard inverted microscope and avoids the requirement for specialized sample mounting procedures.

Sign in to keep reading

3 more chapters are free with a Lably account.

More in Human body