Medical ultrasound

Medical intervention

  • 41 min read
  • 11 chapters
Medical ultrasound
Unknown author · Public domain

Related on Lably

Through Medical diagnosis, Ultrasound and Muscle

Ultrasound is composed of sound waves with frequencies greater than 20,000 Hz, which is the approximate upper threshold of human hearing. Ultrasonic images, also known as sonograms, are created by sending pulses of ultrasound into tissue using a probe. The ultrasound pulses echo off tissues with different reflection properties and are returned to the probe which records and displays them as an image.

A general-purpose ultrasonic transducer may be used for most imaging purposes but some situations may require the use of a specialized transducer. Most ultrasound examination is done using a transducer on the surface of the body, but improved visualization is often possible if a transducer can be placed inside the body. For this purpose, special-use transducers, including transvaginal, endorectal, and transesophageal transducers are commonly employed. At the extreme, very small transducers can be mounted on small diameter catheters and placed within blood vessels to image the walls and disease of those vessels.

Types

Mode

The imaging mode refers to probe and machine settings that result in specific dimensions of the ultrasound image. Several modes of ultrasound are used in medical imaging:

  • A-mode: Amplitude mode refers to the mode in which the amplitude of the transducer voltage is recorded as a function of two-way travel time of an ultrasound pulse. A single pulse is transmitted through the body and scatters back to the same transducer element. The voltage amplitudes recorded correlate linearly to acoustic pressure amplitudes. A-mode is one-dimensional.
  • B-mode: In brightness mode, an array of transducer elements scans a plane through the body resulting in a two-dimensional image. Each pixel value of the image correlates to voltage amplitude registered from the backscattered signal. The dimensions of B-mode images are voltage as a function of angle and two-way time.
  • M-mode: In motion mode, A-mode pulses are emitted in succession. The backscattered signal is converted to lines of bright pixels, whose brightness linearly correlates to backscattered voltage amplitudes. Each next line is plotted adjacent to the previous, resulting in an image that looks like a B-mode image. The M-mode image dimensions are however voltage as a function of two-way time and recording time. This mode is an ultrasound analogy to streak video recording in high-speed photography. As moving tissue transitions produce backscattering, this can be used to determine the displacement of specific organ structures, most commonly the heart.

Most machines convert two-way time to imaging depth using as assumed speed of sound of 1540 m/s. As the actual speed of sound varies greatly in different tissue types, an ultrasound image is therefore not a true tomographic representation of the body.

3D imaging

Three-dimensional imaging is done by combining B-mode images, using dedicated rotating or stationary probes. This has been referred to as C-mode.

A hybrid rotating ultrasound/photoacoustic system produces 3D images, including blood and tissue without the use of magnets or ionizing radiation. It uses an arc of detectors to create a volumetric image. A laser pulse causes hemoglobin molecules to vibrate and generate acoustic signal also absorbed by the detectors.

Technique

An imaging technique is the method of signal generation and processing that supports in a specific application. Most techniques use B-mode.

  • Doppler sonography: This imaging technique makes use of the Doppler effect in detection and measuring moving targets, typically blood.
  • Harmonic imaging: backscattered signal from tissue is filtered to comprise only frequency content of at least twice the centre frequency of the transmitted ultrasound. Harmonic imaging used for perfusion detection when using ultrasound contrast agents and for the detection of tissue harmonics. Common pulse schemes for the creation of harmonic response without the need of real-time Fourier analysis are pulse inversion and power modulation.
  • B-flow is an imaging technique that digitally highlights moving reflectors (mainly red blood cells) while suppressing the signals from the surrounding stationary tissue. It aims to visualize flowing blood and surrounding stationary tissues simultaneously. It is thus an alternative or complement to Doppler ultrasonography in visualizing blood flow.

Therapeutic ultrasound aimed at a specific tumor or calculus is not an imaging mode. However, for positioning a treatment probe to focus on a specific region of interest, A-mode and B-mode are typically used, often during treatment.