Through Voltage, Temperature and Kiln

Commercial thermocouples are inexpensive, interchangeable, are supplied with standard connectors, and can measure a wide range of temperatures. In contrast to most other methods of temperature measurement, thermocouples are self-powered and require no external form of excitation. The main limitation with thermocouples is accuracy; system errors of less than one degree Celsius (°C) can be difficult to achieve.

Thermocouples are widely used in science and industry. Applications include temperature measurement for kilns, gas turbine exhaust, diesel engines, and other industrial processes. Thermocouples are also used in homes, offices and businesses as the temperature sensors in thermostats, and also as flame sensors in safety devices for gas-powered appliances.

Principle of operation

In 1821, the German physicist Thomas Johann Seebeck discovered that a magnetic needle held near a circuit made up of two dissimilar metals got deflected when one of the dissimilar metal junctions was heated. At the time, Seebeck referred to this consequence as thermo-magnetism. The magnetic field he observed was later shown to be due to thermo-electric current. In practical use, the voltage generated at a single junction of two different types of wire is what is of interest as this can be used to measure temperature at very high and low temperatures. The magnitude of the voltage depends on the types of wire being used. Generally, the voltage is in the microvolt range and care must be taken to obtain a usable measurement. Although very little current flows, power can be generated by a single thermocouple junction. Power generation using multiple thermocouples, as in a thermopile, is common.

The standard configuration of a thermocouple is shown in the figure. The dissimilar conductors contact at the measuring (aka hot) junction and at the reference (aka cold) junction. The thermocouple is connected to the electrical system at its reference junction. The figure shows the measuring junction on the left, the reference junction in the middle and represents the rest of the electrical system as a voltage meter on the right.

The temperature Tsense is obtained via a characteristic function E(T) for the type of thermocouple which requires inputs: measured voltage V and reference junction temperature Tref. The solution to the equation E(Tsense) = V + E(Tref) yields Tsense. Sometimes these details are hidden inside a device that packages the reference junction block (with Tref thermometer), voltmeter, and equation solver.

Seebeck effect

The Seebeck effect refers to the development of an electromotive force across two points of an electrically conducting material when there is a temperature difference between those two points. Under open-circuit conditions where there is no internal current flow, the gradient of voltage (∇V) is directly proportional to the gradient in temperature (∇T):

∇V=−S(T)∇T,

where S(T) is a temperature-dependent material property known as the Seebeck coefficient.

The standard measurement configuration shown in the figure shows four temperature regions and thus four voltage contributions:

  1. Change from Tmeter to Tref, in the lower copper wire.
  2. Change from Tref to Tsense, in the alumel wire.
  3. Change from Tsense to Tref, in the chromel wire.
  4. Change from Tref to Tmeter, in the upper copper wire.

The first and fourth contributions cancel out exactly, because these regions involve the same temperature change and an identical material. As a result, Tmeter does not influence the measured voltage. The second and third contributions do not cancel, as they involve different materials.

The measured voltage turns out to be

V=∫TrefTsense(S+(T)−S−(T))dT,

where S+ and S− are the Seebeck coefficients of the conductors attached to the positive and negative terminals of the voltmeter, respectively (chromel and alumel in the figure).

Characteristic function

The thermocouple's behaviour is captured by a characteristic function E(T), which needs only to be consulted at two arguments:

V=E(Tsense)−E(Tref).

In terms of the Seebeck coefficients, the characteristic function is defined by

E(T)=∫TS+(T′)−S−(T′)dT′+const

The constant of integration in this indefinite integral has no significance, but is conventionally chosen such that E(0∘C)=0.

Thermocouple manufacturers and metrology standards organizations such as NIST provide tables of the function E(T) that have been measured and interpolated over a range of temperatures, for particular thermocouple types (see External links section for access to these tables).

Reference junction

To obtain the desired measurement of Tsense, it is not sufficient to just measure V. The temperature at the reference junctions Tref must also be known. Two strategies are often used here:

  • "Ice bath": The reference junction block is maintained at a known temperature as it is immersed in a semi-frozen bath of distilled water at atmospheric pressure. The precise temperature of the melting point phase transition acts as a natural thermostat, fixing Tref to 0 °C.
  • Reference junction sensor (known as "cold junction compensation"): The reference junction block is allowed to vary in temperature, but the temperature is measured at this block using a separate temperature sensor. This secondary measurement is used to compensate for temperature variation at the junction block. The thermocouple junction is often exposed to extreme environments, while the reference junction is often mounted near the instrument's location. Semiconductor thermometer devices are often used in modern thermocouple instruments.

In both cases the value V+E(Tref) is calculated, then the function E(T) is searched for a matching value. The argument where this match occurs is the value of Tsense:

E(Tsense)=V+E(Tref).