Voltage clamp

Method of measuring biological cell currents

  • 14 min read
  • 4 chapters

Related on Lably

Through Experiment, Cell membrane and Neuron

The voltage clamp operates by negative feedback. The membrane potential amplifier measures membrane voltage and sends output to the feedback amplifier; this subtracts the membrane voltage from the command voltage, which it receives from the signal generator. This signal is amplified and output is sent into the axon via the current-passing electrode.

Technique

The voltage clamp is a current generator. Transmembrane voltage is recorded through a "voltage electrode", relative to ground, and a "current electrode" passes current into the cell. The experimenter sets a "holding voltage", or "command potential", and the voltage clamp uses negative feedback to maintain the cell at this voltage. The electrodes are connected to an amplifier, which measures membrane potential and feeds the signal into a feedback amplifier. This amplifier also gets an input from the signal generator that determines the command potential, and it subtracts the membrane potential from the command potential (Vcommand – Vm), magnifies any difference, and sends an output to the current electrode. Whenever the cell deviates from the holding voltage, the operational amplifier generates an "error signal", that is the difference between the command potential and the actual voltage of the cell. The feedback circuit passes current into the cell to reduce the error signal to zero. Thus, the clamp circuit produces a current equal and opposite to the ionic current.